Compare thermocouple types K, J, T, E, N, R, S and B: their metals, ranges, sensitivity and the IEC and ANSI wire colours, and work out the Class 1 and Class 2 tolerance at the temperature you measure.
Thermocouple types compared
| Type | Metals (+ and -) | Usual range | µV per °C | IEC colour | ANSI extension wire | Good for |
|---|---|---|---|---|---|---|
| K | Nickel chromium and nickel aluminium | -200 to 1,260 °C | About 41 | Green | Yellow | General purpose: ovens, kilns, exhausts, most handheld meters and probe inputs |
| J | Iron and copper nickel | -40 to 750 °C | About 52 | Black | Black | Older plant, plastics machinery, vacuum; the iron leg rusts in damp or oxidising air above about 550 °C |
| T | Copper and copper nickel | -200 to 350 °C | About 41 | Brown | Blue | Low temperatures: food, cold chain, freezers, labs; accurate below zero |
| E | Nickel chromium and copper nickel | -200 to 900 °C | About 61 | Violet | Purple | Highest output of the base metal types; low and cryogenic temperatures |
| N | Nickel chromium silicon and nickel silicon | -200 to 1,260 °C | About 27 | Pink | Orange | Like K but drifts less at high temperature; furnaces and heat treatment |
| R | Platinum 13% rhodium and platinum | 0 to 1,600 °C | About 6 (about 13 at 1,000 °C) | Orange | Green | Glass, steel and ceramic furnaces, reference work; expensive and delicate; often rated nearer 1,450 °C for continuous use |
| S | Platinum 10% rhodium and platinum | 0 to 1,600 °C | About 6 (about 11.5 at 1,000 °C) | Orange | Green | Calibration labs and furnaces; long a reference standard; often rated nearer 1,450 °C for continuous use |
| B | Platinum 30% rhodium and platinum 6% rhodium | 600 to 1,700 °C | Near 0 at room temperature, about 9 at 1,000 °C | Check with the maker | Grey | The hottest standard type: steel and glass melting; useless below about 600 °C |
IEC 60584-3: the positive wire and the outer sheath take the type's colour and the negative wire is white. ANSI MC96.1 (common on American imports): the colours shown are for extension wire, where the outer jacket takes the colour and the negative wire is always red; ANSI thermocouple-grade wire for J, K, T, E and N has a brown jacket. We could not confirm an IEC colour for Type B from a published chart, so check it with the maker. Ranges are typical working limits for the wire; the sheath and the atmosphere set the real limit.
How a thermocouple works
Two different metals welded at the tip make a tiny voltage that grows with the difference in temperature between the tip and the other end, the cold junction, where the wires meet the meter. The meter measures the temperature at its own terminals and adds it back (cold junction compensation). That is why the cable from the probe to the meter must be the same type of thermocouple or extension wire: ordinary copper cable or the wrong type of plug creates new junctions and a wrong reading. Mixing up K and J, or the polarity, is a common fault on site.
Choosing a type
- Up to about 350 °C, and below zero: Type T, or a PT100 for better accuracy.
- General work to about 1,100 °C: Type K, which most handheld meters read. Type N where drift at high temperature matters.
- 1,200 to 1,600 °C: Type R or S in a ceramic sheath, or a pyrometer if contact is impractical.
- Above 1,600 °C: Type B, or non-contact.
Put the type, the length, the sheath, the tip (exposed, grounded or insulated), the plug and the number needed on your request; see thermocouples.
Frequently asked questions
What is the most common thermocouple type?
What colour is a Type K thermocouple wire?
Can I use a Type K probe on a Type J meter?
How accurate is a Type K thermocouple?
Other free tools
- Temperature converterCelsius, Fahrenheit, kelvin and Rankine, with oven and reference tables
- Boiling point and thermometer checkWhat water boils at where you are, and how far off your thermometer reads
- Distance to spot calculatorSpot size from the D:S ratio, and how far away you can stand
- Thermal camera FOV and spot sizeField of view, IFOV, pixel size and the smallest spot a camera measures
- Emissivity table and correctionTypical emissivity of common materials, and what a wrong setting does
- Dew point and humidity calculatorDew point from temperature and humidity, and condensation risk on a surface
- PT100 resistance table and calculatorResistance to temperature and back for PT100 and PT1000, with tolerance classes