Look up the typical emissivity of the surface you are measuring, and correct a reading taken at the wrong setting, with the room temperature that the surface reflects taken into account.
What emissivity is
Every surface gives off infrared in proportion to its temperature, but not equally well. Emissivity (ε) is how well, on a scale from 0 to 1: a perfect emitter, a blackbody, is 1. Matt, dark and organic surfaces sit around 0.90 to 0.98, which is why most thermometers come set at 0.95. Shiny metals sit at 0.02 to 0.2: they give off little infrared of their own and reflect the room around them like a mirror, so a gun set at 0.95 reads them well below their true temperature and is easily fooled by your own body heat.
Emissivity table
| Surface | Typical ε | Note |
|---|---|---|
| Human skin | 0.98 | Why forehead thermometers work |
| Water | 0.95 to 0.96 | |
| Ice | 0.96 to 0.98 | |
| Snow | 0.80 to 0.90 | Some tables give up to 0.98 |
| Most food: meat, vegetables, dough | 0.95 to 0.97 | Surface only; use a probe for the core |
| Wood, unfinished | 0.90 to 0.95 | |
| Paper and cardboard | 0.90 to 0.95 | |
| Cloth and textiles | 0.90 to 0.95 | |
| Rubber | 0.90 to 0.95 | |
| Plastics, opaque | 0.85 to 0.95 | Thin film may be see-through |
| Paint, most colours (not metallic) | 0.90 to 0.95 | |
| Matt black paint or electrical tape | 0.95 to 0.97 | Use as a reference patch |
| Concrete | 0.92 to 0.97 | |
| Brick, rough | 0.90 to 0.93 | |
| Asphalt | 0.93 to 0.97 | |
| Tar | 0.79 to 0.84 | |
| Plaster and drywall | 0.85 to 0.95 | |
| Glazed ceramic tiles | 0.90 to 0.94 | |
| Glass, plain sheet | 0.85 to 0.94 | Reads the glass, not what is behind it |
| Soil, dry | 0.90 to 0.95 | |
| Sand | 0.90 | |
| Graphite | 0.70 to 0.98 | Filed or powdered graphite sits near the top |
| Aluminium, polished | 0.02 to 0.10 | Reflects the room: do not trust a reading |
| Aluminium, rough | 0.07 to 0.30 | |
| Aluminium, oxidised | 0.20 to 0.40 | |
| Aluminium, anodised | 0.60 to 0.90 | |
| Stainless steel, polished | 0.10 to 0.20 | |
| Steel, galvanised, new | 0.20 to 0.30 | |
| Steel, oxidised or rusted | 0.60 to 0.90 | Heavy rust is near the top |
| Cast iron, oxidised | 0.60 to 0.90 | |
| Copper, polished | 0.02 to 0.05 | Busbars: use tape or a camera with care |
| Copper, oxidised | 0.60 to 0.80 | |
| Brass, polished | 0.03 to 0.10 | |
| Gold and silver, polished | 0.02 to 0.04 |
Typical values at 8 to 14 µm, the band of most handheld infrared thermometers and thermal cameras, as published in makers' tables. The finish, oxidation, temperature and viewing angle all change the real value; pyrometers working at short wavelengths use their own tables. See pyrometers.
Three ways to get a good reading on a tricky surface
- The tape trick: stick a patch of matt black electrical tape or a dab of matt paint on the surface, wait for it to reach the same temperature, and read the patch at 0.95. Only on a surface that is safe to touch and dead.
- Set ε by comparison: measure a spot with a contact probe or thermocouple, then adjust the emissivity on the infrared thermometer until it agrees. Note the value for that material.
- Read an oxidised or painted part nearby instead of the bright metal, or use a contact probe. A thermometer with adjustable emissivity and a Type K input does both; see temperature guns.
Frequently asked questions
What emissivity should I set on my infrared thermometer?
Why does my temperature gun read too low on stainless steel?
Does emissivity matter for thermal cameras?
My thermometer has fixed emissivity. What can I do?
Other free tools
- Temperature converterCelsius, Fahrenheit, kelvin and Rankine, with oven and reference tables
- Boiling point and thermometer checkWhat water boils at where you are, and how far off your thermometer reads
- Distance to spot calculatorSpot size from the D:S ratio, and how far away you can stand
- Thermal camera FOV and spot sizeField of view, IFOV, pixel size and the smallest spot a camera measures
- Dew point and humidity calculatorDew point from temperature and humidity, and condensation risk on a surface
- PT100 resistance table and calculatorResistance to temperature and back for PT100 and PT1000, with tolerance classes
- Thermocouple types and toleranceK, J, T, E, N, R, S and B: ranges, colour codes and class tolerances