Convert a temperature to the resistance a PT100, PT500 or PT1000 should read, or a measured resistance back to a temperature, with the tolerance of each class and the error the leads add on a 2-wire hookup.
How a PT100 works
A PT100 is a platinum resistance thermometer (RTD) with a resistance of 100 Ω at 0 °C, rising by about 0.385 Ω per degree. A PT1000 is the same with ten times the resistance, so the leads and connections matter ten times less. The relationship is set by IEC 60751 with the Callendar-Van Dusen equation, which is why a PT100 from any maker reads the same on any transmitter or controller set for it. PT100s are more accurate and stable than thermocouples up to a few hundred degrees and are the usual choice for food, pharmaceutical, HVAC and process work; see temperature probes.
Tolerance classes
| Class | Tolerance | At 0 °C | At 100 °C | At 200 °C |
|---|---|---|---|---|
| AA | ± (0.1 + 0.0017 |t|) °C | ± 0.10 °C | ± 0.27 °C | ± 0.44 °C |
| A | ± (0.15 + 0.002 |t|) °C | ± 0.15 °C | ± 0.35 °C | ± 0.55 °C |
| B | ± (0.3 + 0.005 |t|) °C | ± 0.30 °C | ± 0.80 °C | ± 1.30 °C |
Class B is the common standard grade. Class A and AA (also sold as 1/3 DIN, close to AA) cost more and are used where a tight reading matters. The class says how close a new sensor is to the table; a calibration says how close yours is now.
2-wire, 3-wire and 4-wire
On a 2-wire hookup the meter measures the sensor plus both leads, so every ohm of lead resistance reads as about 2.6 °C too hot on a PT100. Long or thin cable makes it worse. A 3-wire hookup cancels most of the lead resistance, and is the industry norm. A 4-wire hookup cancels it fully and is used in labs and for reference thermometers. A PT1000 on 2 wires suffers a tenth of the error, which is why it suits battery loggers and HVAC controllers.
PT100 and PT1000 resistance table
| Temperature | PT100 Ω | PT1000 Ω |
|---|---|---|
| -50 °C | 80.31 | 803.1 |
| -40 °C | 84.27 | 842.7 |
| -30 °C | 88.22 | 882.2 |
| -20 °C | 92.16 | 921.6 |
| -10 °C | 96.09 | 960.9 |
| 0 °C | 100.00 | 1,000.0 |
| 10 °C | 103.90 | 1,039.0 |
| 20 °C | 107.79 | 1,077.9 |
| 30 °C | 111.67 | 1,116.7 |
| 40 °C | 115.54 | 1,155.4 |
| 50 °C | 119.40 | 1,194.0 |
| 60 °C | 123.24 | 1,232.4 |
| 70 °C | 127.08 | 1,270.8 |
| 80 °C | 130.90 | 1,309.0 |
| 90 °C | 134.71 | 1,347.1 |
| 100 °C | 138.51 | 1,385.1 |
| 110 °C | 142.29 | 1,422.9 |
| 120 °C | 146.07 | 1,460.7 |
| 130 °C | 149.83 | 1,498.3 |
| 140 °C | 153.58 | 1,535.8 |
| 150 °C | 157.33 | 1,573.3 |
| 160 °C | 161.05 | 1,610.5 |
| 170 °C | 164.77 | 1,647.7 |
| 180 °C | 168.48 | 1,684.8 |
| 190 °C | 172.17 | 1,721.7 |
| 200 °C | 175.86 | 1,758.6 |
| 210 °C | 179.53 | 1,795.3 |
| 220 °C | 183.19 | 1,831.9 |
| 230 °C | 186.84 | 1,868.4 |
| 240 °C | 190.47 | 1,904.7 |
| 250 °C | 194.10 | 1,941.0 |
| 260 °C | 197.71 | 1,977.1 |
| 270 °C | 201.31 | 2,013.1 |
| 280 °C | 204.90 | 2,049.0 |
| 290 °C | 208.48 | 2,084.8 |
| 300 °C | 212.05 | 2,120.5 |
| 310 °C | 215.61 | 2,156.1 |
| 320 °C | 219.15 | 2,191.5 |
| 330 °C | 222.68 | 2,226.8 |
| 340 °C | 226.21 | 2,262.1 |
| 350 °C | 229.72 | 2,297.2 |
| 360 °C | 233.21 | 2,332.1 |
| 370 °C | 236.70 | 2,367.0 |
| 380 °C | 240.18 | 2,401.8 |
| 390 °C | 243.64 | 2,436.4 |
| 400 °C | 247.09 | 2,470.9 |
| 410 °C | 250.53 | 2,505.3 |
| 420 °C | 253.96 | 2,539.6 |
| 430 °C | 257.38 | 2,573.8 |
| 440 °C | 260.78 | 2,607.8 |
| 450 °C | 264.18 | 2,641.8 |
| 460 °C | 267.56 | 2,675.6 |
| 470 °C | 270.93 | 2,709.3 |
| 480 °C | 274.29 | 2,742.9 |
| 490 °C | 277.64 | 2,776.4 |
| 500 °C | 280.98 | 2,809.8 |
Frequently asked questions
What is the resistance of a PT100 at 100 °C?
How do I test a PT100 with a multimeter?
PT100 or thermocouple?
What does alpha 0.00385 mean?
Other free tools
- Temperature converterCelsius, Fahrenheit, kelvin and Rankine, with oven and reference tables
- Boiling point and thermometer checkWhat water boils at where you are, and how far off your thermometer reads
- Distance to spot calculatorSpot size from the D:S ratio, and how far away you can stand
- Thermal camera FOV and spot sizeField of view, IFOV, pixel size and the smallest spot a camera measures
- Emissivity table and correctionTypical emissivity of common materials, and what a wrong setting does
- Dew point and humidity calculatorDew point from temperature and humidity, and condensation risk on a surface
- Thermocouple types and toleranceK, J, T, E, N, R, S and B: ranges, colour codes and class tolerances