PT100 and PT1000 resistance table and calculator

A PT100 reads 100 Ω at 0 °C and about 0.385 Ω more for every degree. Type a temperature or a resistance to convert it, see the tolerance of each class and what the leads add on a 2-wire hookup.

PT100 resistance table and calculatorCallendar-Van Dusen equation, IEC 60751
Updates
AS YOU TYPE
Convert
Resistance
-
Sensitivity
-
Class AA tolerance
-
Class A tolerance
-
Class B tolerance
-
2-wire lead error
-

R = R0 (1 + A t + B t² + C (t - 100) t³), with C used only below 0 °C. Tolerance classes from IEC 60751; each class applies over its own temperature range, which depends on the sensor build.

Convert a temperature to the resistance a PT100, PT500 or PT1000 should read, or a measured resistance back to a temperature, with the tolerance of each class and the error the leads add on a 2-wire hookup.

How a PT100 works

A PT100 is a platinum resistance thermometer (RTD) with a resistance of 100 Ω at 0 °C, rising by about 0.385 Ω per degree. A PT1000 is the same with ten times the resistance, so the leads and connections matter ten times less. The relationship is set by IEC 60751 with the Callendar-Van Dusen equation, which is why a PT100 from any maker reads the same on any transmitter or controller set for it. PT100s are more accurate and stable than thermocouples up to a few hundred degrees and are the usual choice for food, pharmaceutical, HVAC and process work; see temperature probes.

Tolerance classes

ClassToleranceAt 0 °CAt 100 °CAt 200 °C
AA± (0.1 + 0.0017 |t|) °C± 0.10 °C± 0.27 °C± 0.44 °C
A± (0.15 + 0.002 |t|) °C± 0.15 °C± 0.35 °C± 0.55 °C
B± (0.3 + 0.005 |t|) °C± 0.30 °C± 0.80 °C± 1.30 °C

Class B is the common standard grade. Class A and AA (also sold as 1/3 DIN, close to AA) cost more and are used where a tight reading matters. The class says how close a new sensor is to the table; a calibration says how close yours is now.

2-wire, 3-wire and 4-wire

On a 2-wire hookup the meter measures the sensor plus both leads, so every ohm of lead resistance reads as about 2.6 °C too hot on a PT100. Long or thin cable makes it worse. A 3-wire hookup cancels most of the lead resistance, and is the industry norm. A 4-wire hookup cancels it fully and is used in labs and for reference thermometers. A PT1000 on 2 wires suffers a tenth of the error, which is why it suits battery loggers and HVAC controllers.

PT100 and PT1000 resistance table

TemperaturePT100 ΩPT1000 Ω
-50 °C80.31803.1
-40 °C84.27842.7
-30 °C88.22882.2
-20 °C92.16921.6
-10 °C96.09960.9
0 °C100.001,000.0
10 °C103.901,039.0
20 °C107.791,077.9
30 °C111.671,116.7
40 °C115.541,155.4
50 °C119.401,194.0
60 °C123.241,232.4
70 °C127.081,270.8
80 °C130.901,309.0
90 °C134.711,347.1
100 °C138.511,385.1
110 °C142.291,422.9
120 °C146.071,460.7
130 °C149.831,498.3
140 °C153.581,535.8
150 °C157.331,573.3
160 °C161.051,610.5
170 °C164.771,647.7
180 °C168.481,684.8
190 °C172.171,721.7
200 °C175.861,758.6
210 °C179.531,795.3
220 °C183.191,831.9
230 °C186.841,868.4
240 °C190.471,904.7
250 °C194.101,941.0
260 °C197.711,977.1
270 °C201.312,013.1
280 °C204.902,049.0
290 °C208.482,084.8
300 °C212.052,120.5
310 °C215.612,156.1
320 °C219.152,191.5
330 °C222.682,226.8
340 °C226.212,262.1
350 °C229.722,297.2
360 °C233.212,332.1
370 °C236.702,367.0
380 °C240.182,401.8
390 °C243.642,436.4
400 °C247.092,470.9
410 °C250.532,505.3
420 °C253.962,539.6
430 °C257.382,573.8
440 °C260.782,607.8
450 °C264.182,641.8
460 °C267.562,675.6
470 °C270.932,709.3
480 °C274.292,742.9
490 °C277.642,776.4
500 °C280.982,809.8

Frequently asked questions

What is the resistance of a PT100 at 100 °C?
138.51 Ω. At 0 °C it is 100.00 Ω, at 50 °C 119.40 Ω and at 200 °C 175.86 Ω.
How do I test a PT100 with a multimeter?
Disconnect it, measure across the element on the ohms range and compare with the table at the known temperature: about 108 Ω at room temperature (21 °C). Short the meter leads first and subtract their resistance. An open circuit or a reading far off means a broken or damaged sensor.
PT100 or thermocouple?
A PT100 is more accurate and stable from about -200 to 600 °C. A thermocouple is cheaper, tougher, faster and goes much hotter, to 1,200 °C or more. See thermocouple types.
What does alpha 0.00385 mean?
The average change in resistance per degree between 0 and 100 °C, divided by the resistance at 0 °C. 0.00385 is the IEC 60751 standard; some older or American sensors use 0.00392, so check before swapping.

Other free tools

Free quote request, no obligation

Request quotes for temperature probes and PT100 sensors

Describe it the way a technician would: how far away you stand and how big the target is, or the camera and the distance, or simply how many. The planner draws the measuring cone or the camera's view as you type. Suppliers and service companies in your area quote you directly.

Measurement sheet and quote requestTemperature Probes and PT100 Sensors
Req
40611
Date
2026-11-20
Cost
FREE

1The job

2How many (optional, it helps the quote)

YOUR REQUESTROUGH, NOT TO SCALE

-
-
-
-
-
-
Estimate
-

Plain geometry for the request, not a measurement.

3Where you are now

4Your list (optional)

  1. 01
  2. 02
Up to 12 lines: instruments, probes, certificates or services.
Needs and options

5Your details

Free and no obligation. By sending you agree that we may share your request with suitable suppliers and service companies. See our privacy policy.

Get a free quote